My SciELO
Services on Demand
Journal
Article
Indicators
- Cited by SciELO
Related links
- Similars in SciELO
Share
Nucleus
Print version ISSN 0864-084XOn-line version ISSN 2075-5635
Abstract
LOPEZ DIAZ, Adlín et al. Combined parameters estimation for dosimetry treatment planning in Hyperthyroidism using 2D and 3D. Nucleus [online]. 2017, n.61, pp.16-20. ISSN 0864-084X.
Optimization and verification of Specific Patient Treatment Planning with unsealed sources in hyperthyroidism diseases is a desirable goal from medical and radiation protection viewpoints. In order to verify the estimation of patient’s specific treatment dose and his/her related parameters, a combination of 3 different apparatus or pieces of equipment used in nuclear medicine were studied - the Iodine Probe, a Philips Forte Camera with pin-hole collimators and a Mediso Nucline with HEGP for planar and SPECT techniques- by using the typical neck phantom and sources simulating diagnosis and treatment procedure. The linear behavior on diagnostic and therapeutic activity range was verified, showing a linear correlation fitting factor > 0,99. The differences between thyroid uptake determinations in all equipment were less than 6 % for therapeutic activities and less than 1,1 % in the diagnostic range. The combined protocol to calculate all the necessary parameters for the patient treatment dose planning using 2D or 3D approach was established and verified, avoiding wasting time with gamma cameras and with only one administration of
Keywords : optimization; dosimetry; patients; theraphy; iodine 131; hyperthyroidism; verification.