SUAREZ RODRIGUEZ, Yeila; PRIETO MIRANDA, Enrique Francisco; CHAVEZ ARDANZA, Armando y BARRERAS GONZALEZ, Gisela. Methodology for validation and control of irradiation process in a self-shielding gamma irradiator. Nucleus [online]. 2017, n.61, pp. 44-47. ISSN 2075-5635.
A methodology was developed for the validation and routine control of the irradiation process performed in a self-shielding gamma irradiator with 60Co sources. This methodology was applied to an irradiator model ISOGAMMA LLCo in order to ensure the proper operation of the equipment, efficiency of services provided and to consistently obtain, , a product that meets the quality attributes and default specifications
Palabras clave : validation; irradiation process; self-shielding gamma irradiator; absorbed dose; dose rate.