SciELO - Scientific Electronic Library Online

 
vol.35 issue1Strontium stannate as an alternative anode for alkali-ion batteriesGd2O3 nanoparticles grafted to Amylovis®-204, with potential affinity for β-amyloid plaques. Part II author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Services on Demand

Article

Indicators

  • Have no cited articlesCited by SciELO

Related links

  • Have no similar articlesSimilars in SciELO

Share


Revista Cubana de Química

On-line version ISSN 2224-5421

Abstract

QUINONES-GARCIA, Yaidel; CARABALLOSO-NOA, Isabel; CALVO-ALONSO, Amalia  and  TORRE-LOPEZ, Jorge de la. Uncertainty estimation in the measurement of furvina content in Vitrofural®. Rev Cub Quim [online]. 2023, vol.35, n.1, pp. 14-28.  Epub Jan 17, 2023. ISSN 2224-5421.

All results of analytical measurements have a degree of uncertainty, which is related to the steps carried out and the materials and/or equipment used in the method. The objective of this work was to estimate the uncertainty in the measurement of the Furvina content in Vitrofural by ultraviolet - visible spectrophotometry, through the GUM approach. The mathematical model was obtained and the sources of uncertainty were identified with a cause-effect diagram. Uncertainty components were evaluated and the combined uncertainty value (0,06 %) was obtained. The greatest contributions to uncertainty correspond to glassware used. The expanded uncertainty (0,12 %) was calculated and the Furvina content was determined in a batch of Vitrofural: the final result was 30,21 % ± 0,12 %, k = 2.

Keywords : uncertainty; Vitrofural; furvina; ultraviolet-visible spectrophotometry..

        · abstract in Spanish     · text in Spanish     · Spanish ( pdf )