SciELO - Scientific Electronic Library Online

 
 issue55Monitoring intervals for the measurement of the radionuclides 125I and 129I in the thyroid glandRadiological safety assessment in the practice of well logging using risk matrixes author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Services on Demand

Article

Indicators

  • Have no cited articlesCited by SciELO

Related links

  • Have no similar articlesSimilars in SciELO

Share


Nucleus

Print version ISSN 0864-084X

Abstract

ORAMAS POLO, Ivón  and  FIGUEROA DEL VALLE, Diana G. Risk analysis for working with an analyzer for gamma cameras diagnostic. Nucleus [online]. 2014, n.55, pp. 19-23. ISSN 0864-084X.

In this work, a risk analysis for working with an analyzer for gamma cameras diagnostic was made. The method employed is based on determining the Hazard Rating Number (HRN). The results showed that the risks with higher value of HRN are electrocution with 100 and touch source container with hands with 75. These risks were classified as "Very High" and "High" respectively. The following risks were classified as "Important":  Fall of the source container (HRN = 25), high dose of the sample in the container (HRN = 20) and fracture of glass detector (HRN = 30). The wrong shielding of the source container (HRN= 10) is a risk that was classified as "Low". Safety rules for use of the system are indicated. An action plan for risk management is also presented

Keywords : hazards; photomultipliers; gamma cameras; risk assessment.